The K850 Critical Point Dryer features thermo-electronic heating and adiabatic cooling, allowing precise temperature control between +5°C and +35°C. This ensures accurate attainment of the critical point by pre-cooling the chamber. It has three valves: a fluid inlet, a flushing valve, and a fine needle valve for controlled gas venting and pressure release. Additionally, a built-in magnetic stirrer ensures thorough mixing of specimens with circulating fluids.
Automatic thermoelectric heating - Adiabatic cooling - Digital temperature monitor - Dimensions and weight: 450 mm w x 350 mm d x 175 mm h, 12 kg -
Recipient: 30 mm Ø x 40 mm h, vertically mounted with CO2-inlet at top and drain of solvent at bottom - Fitted with thermo-electric heating and adiabatic cooling enabling temperature control of +5 °C cooling and +35 °C during heating, thermometer (0 – 120 °C) und manometer (0 – 2.000 p.s.i./ca. 137 bar), normal working temperature and pressure 35 °C @ 1.350 p.s.i. (approx. 93 bar) - Fitted with three valves: fluid inlet, flushing and a gas venting system which uses a fine needle valve to give controlled pressure let down - Magnetic stirrer for enhanced solvent exchange - Side viewing port